INTRODUCTION

 

In Germany, emergency departments (EDs) often suffer from overcrowding, which may lead to stressful working conditions, increased risks for patients and costs for additional staff, as well as for investments in equipment. Furthermore, overcrowding has the potential to extend patient waiting times, in some cases rising up to 6 hours and more.1

 

Over the last few years, an increase of patient demand in hospital EDs by at least 7–10% has been observed.1 This is likely to be partly stimulated by the increasing numbers of elderly people found in Europe, including Germany, which is expected to increase significantly from 2010 onwards.2

 

Limited resources may also cause additional burdens. While government health spending in Europe tended to be maintained at the start of the economic crisis, cuts in spending really began to take effect in 2010 in response to budgetary pressures and the need to reduce large deficits and debts.2 Furthermore, between 2000 and 2010, a total of 245 hospitals closed in Germany, and there was an 11% decrease in hospital beds.2

 

Addressing the efficiency of patient testing might be a solution to problems faced in EDs. Average length of stay (ALOS) for patients is often dependent on the turn-around-time (TAT) of crucial laboratory tests. The laboratory process, therefore, has the potential to strongly influence patient throughput in the ED.

 

Point of care testing (POCT) has been shown in some scenarios to increase the number of patients discharged in a timely manner, determine the priority of urgent but nonemergency patients and decrease delays to treatment initiation.3 The primary advantages provided by POCT devices are increased portability and speed. Using POCT, caregivers can perform, analyse, obtain and act on test results at the bedside in a matter of minutes. Used effectively, POCT has the potential to alleviate the negative effects of overcrowding.1,3-5

 

This white paper describes the results of a randomised, single-centre trial looking at POCT efficacy within the setting of the busy ED at Augsburg Hospital in Germany. In addition, a survey of employee and patient satisfaction with POCT was conducted after implementation.

 

OVERVIEW AND RATIONALE

 

Located in Augsburg, Germany, the Augsburg Hospital is a teaching hospital for the Ludwig-Maximilians-University in Munich. The hospital consists of 25 clinics and institutes and houses more than 5,400 employees, including specialists and supportive personnel.6,7

 

The Augsburg hospital treats approximately 200 patients in its ED each day; this translates to nearly 70,000 patients per year. The ED suffers from a number of resource limitations: physicians, nurses, numbers of examination and treatment rooms; this allows a maximum capacity of only five patients. Overcrowding occurs when the admission rate of emergency patients is higher than the discharge rate of comprehensively diagnosed/treated patients. When this occurs, the amount of patients needing to be treated at any one time is much greater than the ED’s maximum capacity. Effects of overcrowding at the Augsburg Hospital ED are particularly apparent during the morning and in the early evening. The consequence of crowding effects from the patient’s perspective is described by Anantharaman and Seth.8

 

Approximately one-fifth of ED patients visiting Augsburg Hospital suffer from non-specific chest pain. In order to determine if these patients have acute coronary syndrome (ACS), a troponin test (cTn) is performed. Troponin is a protein that is released when heart muscle is damaged, and higher than normal levels may indicate ACS.9 Most (>70%) potential ACS patients within the Augsburg Hospital ED need additional cTn testing in order to confirm their diagnosis. As a result, an estimated 9,000 to 12,000 troponin tests are performed per year.

 

The normal TAT for troponin tests can vary between 42 and 121 minutes when performed by the Augsburg Hospital Central Laboratory. In terms of “six sigma methodology” the extreme variation is a sign of an uncontrolled service.10,11 This can lead to an extended LOS, increased waiting times for patients, stressful working situations and higher costs caused by additional staff and overtime payments. Therefore, it was hypothesised that POCT for troponin would reduce ALOS for ACS patients and thereby lead to economic benefits.Furthermore, it was speculated that patients with ACS would benefit clinically from the quick results provided by the Contemporary Troponin-I (cTnI) test that is available on the i-STAT System, a POCT device.12

 

THE i-STAT SYSTEM

 

In the clinical trial, the i-STAT System from Abbott Point of Care was chosen as the POCT device to be tested. The i-STAT System provides lab-quality test results within ten minutes using only small amounts of blood. The i-STAT cTnI test cartridge uses a two-site enzyme-linked immunosorbant assay (ELISA) method, which culminates in the release of an electrochemically detectable product. Sensors measure the product, which is proportional to the concentration of the cTnI within the sample.12 The reasons for choosing the i-STAT System included:

 

  • Lab-quality, precise results
  • Short TAT for tests
  • Ease of use
  • Ease of integration in the Laboratory Information System (LIS) and Hospital Information System (HIS) of Augsburg Hospital
  • Ease of compliance with the Guidelines of the German Federal Medical Council (RiliBÄK) on quality control checks and documentation13,14

 

METHODS

 

Patients

 

Patients with suspicion of Non-ST-Segment elevation (NSTE)-ACS were investigated. Data was included from those with or without a second troponin test (in Germany, serial testing is not always performed). Patients were excluded if they were treated outside the confines of the ED, or if they were transferred to the ICU before the results of the test were available.

 

Study design

 

A randomised, single-centre trial (see Figure 1) was conducted at the Augsburg Hospital ED. The TATs of troponin tests were compared between the POCT setting and Central Lab setting in a series of three phases.

 

In the first study phase, troponin measurement of patients was performed in the Central Lab setting only. During the first week after having implemented a POCT solution for troponin, the second phase was initiated. In this phase, patients were tested in parallel for troponin in the Central Lab and through POCT. Six months later, the third phase took place, and POCT was used exclusively for patient testing.

 

Troponin tests
 

During the first and the second phase of the study, the Central Lab used a high-sensitive Troponin-T (hsTnT) test with a cut-off of 0.040 ng/mL. The POCT device used a Contemporary Troponin-I test (cTnI) with a cut-off of 0.08 ng/mL throughout the study.

 

Points of measurement

 

The primary outcome measure was TAT (defined as time from blood draw to test result). Additionally therapeutic TAT (tTAT; defined as time from test result to therapeutic decision) and LOS were measured. In order to determine employee satisfaction and acceptance of POCT technology, ED-staff who performed POCT (26 people) were queried on different factors (e.g. satisfaction with workflow effectiveness and patient risks). Patient satisfaction was also measured using a survey. Furthermore, the economic relevance of a POCT investment was calculated and investigated in combination with a resource scheduling and smoothing approach and a queuing theory application, both specifically tailored to the Augsburg data setting.15

 

Economic benefit calculations

 

In this study, reductions in staff costs were determined by estimating reductions in ACS patient monitoring that were the result of a reduced ALOS.

 

  • Staff cost savings = observation cycle (4 times) x observation time per round (3 minutes) x nurse costs (¤0.48 per minute) = ¤5.76/patient In the hospital setting, it is believed infrastructure investments were avoided through implementation of the i-STAT system. This was determined by estimating the savings incurred by not having to install an additional examination room, which would service an approximate 5,730 patients/room/year. Additional rooms cost between ¤110,000 and ¤240,000 to staff/year and require a one-time investment of ¤28,000 for equipment.
  • Reduction in patient density/year = patient density reduction/day x 365 days/year
  • Staff cost savings = [(reduction in patient density/year)/(5,730 patients/room/year)]x ¤110,000-240,000/room/year
  • Equipment cost savings = ¤28,000/room

 

RESULTS

 

The i-STAT System provides clinical benefits in an ED setting

The i-STAT System accelerated the availability of cTn test results and allowed physicians to be notified more quickly

 

The i-STAT System cTnI results were on average available in 14 minutes vs. 70 minutes for the Central Lab, which translates to an improvement of 80% (Figure 2A). Moreover, there was a reduction in the time to physician’s notification of results: 22 minutes with the i-STAT System vs. 92 minutes on average for the Central Lab results (Figure 2B).

 

The i-STAT System reduced the time to delivery of test results needed to make clinical decisions in the ED and led, as a consequence in this study, to a shorter ALOS

 

With the i-STAT System, the time to delivery of results for use in clinical decisions was 40 minutes, which was significantly shortened in comparison to average tTAT for results coming from the Central Lab (110 minutes). This improvement led to an average reduction of ALOS of 70 minutes (Figure 3).

 


Patient density is estimated to decrease in the ED after i-STAT implementation

 

Resources available to the ED limit the maximum number of patients that can be treated at any one time to five. During peak moments, patient demand exceeds ED resources, and patients need to wait.

 

To determine the effects of reduced ALOS on patient density during a typical day, a patient density model based on an assumption of 185 patients per day was utilised with the ALOSs obtained during the trial. The following outcomes were found:

 

• Phase 1 (Central Lab): patient density was nearly 40 patients during peak hours (Figure 4)

• Phase 3 (POCT setting): patient density was reduced during peak hours to 33 patients (Figure 4)

 

According to the capacity utilisation model (i.e. calculation) in the ED, with the reduced ALOS from POCT, the required capacity could be reduced by an equivalent of 7 patients.

 


The i-STAT System provides economic benefits in an ED setting

 

Implementation of the i-STAT System has the potential to reduce staff costs

Using information collected from the study on the time needed for patient observation and an ALOS reduction of 70–90 minutes, it was estimated that the i-STAT System saved approximately ¤5.76 per ACS patient in staff costs (reduction of patient monitoring time). Based on 30 ACS patients per day, this totalled ¤179 per day and ¤62,208 per year.

 


Costs for an additional examination room were avoided through use of the i-STAT System

 

The reduction of ED overcrowding by lowering the patient density by approximately 7 patients (between 9 am and 6 pm) translates to an estimated reduced patient density of 2,555 patients per year. In effect, this would allow the ED to avoid building an additional examination room. It was calculated that POCT would save 45% of staff costs (¤110,000) associated with the new room and the cost of additional examination room equipment (¤28,000). This is a total value of ¤138,000 in avoided investments.

 

The economic benefits incurred through the i-STAT System outweigh the costs of implementing the device

 

The calculated investment for the i-STAT System was ¤3,730 per year, which is based on the depreciation of the devices as well as the required, weekly quality controls. Compared to the economic benefits, investment in the i-STAT System, even when used only for troponin tests of ACS patients, pays off economically.

 

Risk-weighted process analysis, a graphical illustration of the hypothesis, demonstrated POCT is advantageous compared to Central Lab testing in terms of quality, risk and cost

 

Risk-weighted process analysis (RPA) is a powerful Health Technology Assessment (HTA) approach that compares alternative technological/organisational settings in terms of quality, risk and costs.15 Information inputted included:

 

• Quality: Medical quality and outcome

• Cost: ALOS, opportunity costs and lifecycle/operating costs

• Risk: Mortality, acute myocardial infarction

• Settings: Alternative approaches, price and investment costs

 

The resulting RPA is depicted in Figure 5.

 


Staff and patient satisfaction

 

The i-STAT System improved employee satisfaction with the ED process

 

A survey that was conducted revealed that the majority of staff were highly satisfied with process effectiveness and efficiency associated with the i-STAT System implementation in Phase 3 as compared to the Central Lab setting in Phase 1 (Figure 6). Although the majority of the staff were not yet highly satisfied with the process organisation in Phase 3, satisfaction in this category was enormously improved relative to Phase 1 (0% highly satisfied; Figure 6).

 


Patient satisfaction improved

Overall patient satisfaction was increased due to reduced ED waiting times.

 

DISCUSSION

 

Use of POCT devices within an ED setting has been shown to increase the number of patients discharged in a timely manner, determine the priority of urgent patients and decrease delays to treatment initiation.3 Within the setting of the Augsburg Hospital ED, POCT contributed to reducing “crowding effects”, containing process costs and increasing patient satisfaction because of reduced ED waiting times. Furthermore, it increased the satisfaction of the ED staff with the ED process.

 

One reason that the staff preferred the POCT setting is the clinical relevance of POCT for ACS patients. POCT offers the opportunity for faster results, so diagnosis can be made more quickly, using Contemporary cTnI tests. Measurement of cTn in the POCT setting was also significantly more acceptable to the ED staff than testing in the Central Lab setting because POCT is considered more of a self-contained process.

 

Since implementation of the i-STAT System, it has been recognised that a change of setting from Central Lab testing to POCT, in fact, means a shift of workload from Central Lab to the ED staff. The employees’ motivation to use a newly implemented technology, as a part of a new workflow organisation, is crucial to achieving a high level of effectiveness and efficiency. For this reason, it is recommended that the Central Lab be given the task of quality control for POCT devices in order to reduce the burden on ED staff from non-clinical duties.

 

After performing a confounding analysis on data from the study, it was found that a dysfunctional organisation might overrule positive effects of a POCT solution. The delay between the availability of the test result and notification of the attending physician was found to be particularly sensitive to poor organisational workflow and could still lead to increased ALOS of patients.5 In order to continue profiting from a POCT setting, organisational workflow will need to be addressed within the Augsburg Hospital ED.

 

Due to the success of the i-STAT System with troponin testing, it was recommended that additional parameters should be leveraged within the ED department. Future candidate tests include C-reactive protein, creatine kinase-MB, and tests useful for determining the speed of blood coagulation, such as the prothrombin time test (PT/INR). All of these tests, with exception of C-reactive protein, can be measured with the POCT device used in the study: the i-STAT System.

 

CONCLUSIONS

 

Within the setting of the Augsburg Hospital ED, implementation of the i-STAT System contributed to:

 

• Reducing the effects of overcrowding

• Containing process costs

• Increasing patient satisfaction due to reduced ED waiting times

• Increasing ED staff satisfaction with the ED process

 

Source & Image Credit: Abbott

 

References

 

  1. von Eiff MC and von Eiff W. Role and Function of the Emergency Department in a Boundaryless Hospital. Optimizing the Process Flow of Acute Coronary Syndrome Patients under life-threatening Conditions. In: Albach et al (eds.). The Boundaryless Hospital. Berlin, Heidelberg, 2016.
  2. Hope: European Hospital and Healthcare Federation. Hospitals in Europe Healthcare Data 2012. Available at: http://www.hope.be/03activities/quality_eu-hospitals/eu_country_profiles/00-hospitals_in_europe-synthesis.pdf [Accessed March 2016].
  3. Rooney KD and Schilling UM. Point-of-care testing in the overcrowded emergency department – can it make a difference? Critical Care 2014;18:692.
  4. Holland LL et al. Reducing Laboratory Turnaround Time Outliers Can Reduce Emergency Department Patient Length of Stay – An 11-Hospital Study. Am J Clin Pathol 2005;124:672–674.
  5. Renaud B et al. Impact of point-of-care testing in the emergency department evaluation and treatment of patients with suspected acute coronary syndromes. Acad Emerg Med 2008;15(3):216–224.
  6. Klinikum Augsburg. Über uns. Available at: http://www.klinikum-augsburg.de/6/Ueber_uns.htm [Accessed March 2016].
  7. Klinikum Augsburg. Clinic for Heart and Thoracic Surgery. Available at: http://www.klinikum-augsburg.de/4617/Medical_Services/Medical_Surgical_Departments/Clinic_for_Heart_and_Thoracic_Surgery.htm [Accessed March 2016].
  8.  Anantharaman V and Seth P. Emergency Department Overcrowding. In: Kayden S et al (eds.) Emergency Department Leadership and Management. Cambridge 2015:257–269.
  9. MedLinePlus. Troponin test. Available at: https://www.nlm.nih.gov/medlineplus/ency/article/007452.htm[Accessed March 2016].
  10. Buttigieg SC et al. Process Optimization in the Emergency Department by the use of POCT in life-threatening Conditions: Comparative Best Practice Example from Germany and Malta. In: Buttigieg C et al. (eds). International Best Practices in Health Care Management, Vol. 17. Emerald Bingley (UK) 2015:195–219.
  11. von Eiff MC. Patient Risks and Avoidable Costs: Six Sigma Reduces Variation in Medicine. Hospital 2007;9(5):22–24.Available at: https://healthmanagement.org/c/hospital/issuearticle/patient-risks-and-avoidable-costs-six-sigma-reducesvariation-in-medicine [Accessed March 2016].
  12. Abbott Point of Care Inc. Cardiac Troponin I/(cTnI). 2013. Available at: https://www.abbottpointofcare.com/download?docUri=/static-assets/technical-documentation/715595-00U.pdf [Accessed March 2016].
  13. German Medical Association (BAEK). Revision of the “Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations – Rili-BAEK” (unauthorized translation). J Lab Med 2015;39:26–69.
  14.  Abbott Point of Care. i-STAT® 1 System Manuel. Rev. Date: 07-Mar-13. Available at: https://www.abbottpointofcare.com/download?docUri=/static-assets/technical-documentation/i-STAT_1_System_Manual_US_English_014331-00_45A.pdf [Accessed March 2016].
  15. de Bruin AM et al. Modeling the Emergency Cardiac Inpatient Flow: An Application of Queuing Theory. Health Care Management Science 2007;10: 125–137

 